Liquichek Unassayed Chemistry Control

Recalling firm
Bio-Rad New Zealand
Product
Liquichek Unassayed Chemistry Control. [Device]. Affected: Lot Number: 93031, 93041, 93032, 93042, 93030, 93040; Lot Numbers: 56761, 56762, 56763, 56760; Lot Numbers: 46021, 46041, 46022, 46042, 46023, 46043, 46020, 46040; Lot Numbers: 86991, 86992, 86993, 86990. Also covers: Liquid Unassayed Multiqual; Liquid Assayed Multiqual; Liquid Assayed Multiqual Premium
Reason
The manufacturer has identified a stability issue affecting the quality control material lots listed above. Investigation found that Triglycerides, Phosphorus, and Magnesium values increase over time after the vial is thawed and opened (open-vial stability). Magnesium values also decrease over the product's shelf life., Use of affected quality control lots may result in QC shifts for the impacted analytes, which could require laboratory investigation, repeat testing, or use of alternate control material with subsequent temporary operational delays related to QC investigation. Laboratories...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-02
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/ffaead59-594a-4525-93ba-430e8c62522d