Liquichek Unassayed Chemistry Control
- Recalling firm
- Bio-Rad New Zealand
- Product
- Liquichek Unassayed Chemistry Control. [Device]. Affected: Lot Number: 93031, 93041, 93032, 93042, 93030, 93040; Lot Numbers: 56761, 56762, 56763, 56760; Lot Numbers: 46021, 46041, 46022, 46042, 46023, 46043, 46020, 46040; Lot Numbers: 86991, 86992, 86993, 86990. Also covers: Liquid Unassayed Multiqual; Liquid Assayed Multiqual; Liquid Assayed Multiqual Premium
- Reason
- The manufacturer has identified a stability issue affecting the quality control material lots listed above. Investigation found that Triglycerides, Phosphorus, and Magnesium values increase over time after the vial is thawed and opened (open-vial stability). Magnesium values also decrease over the product's shelf life., Use of affected quality control lots may result in QC shifts for the impacted analytes, which could require laboratory investigation, repeat testing, or use of alternate control material with subsequent temporary operational delays related to QC investigation. Laboratories...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-11-02
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory