Generic Name: General-purpose Diagnostic Imaging System Workstation Trade Name: Xyostation Revolus Rl
Summary Xyosoft Corporation is recalling 76 Xyostation Revolus RL diagnostic imaging workstations manufactured between March 2023 and January 2026 due to a software bug that resets manually corrected heart valve analysis measurements back to automatic values after certain operations. While this could potentially lead to incorrect diagnoses or treatment, the company states that doctors typically use multiple clinical indicators to confirm diagnoses and no patient injuries have been reported.
- Recalling firm
- Name of manufacturer and distributor: Xyosoft Corporation Location of manufacturer and distributor: Mita Kokusai Building, 4-28, Mita 1-chome, Minato-ku, Tokyo Type of license: Type 1 medical...
- Product
- Lot: Ser.No:以下の通り RL230005A, RL230028A, RL230046S, RL230055A, RL230027A, RL230071A, RL230094S, RL230107A, RL230116S, RL230118S, RL230119S, RL230128A, RL230172S, RL230182A, RL240009S, RL240017A, RL230181A, RL240041A, RL240063A, RL240064A, RL240077S, RL240104A, RL240108A, RL240111A, RL240100A, RL240129S, RL240110A, RL240115A, RL240132A, RL240148S, RL240119A, RL240151A, RL240170S, RL240147S, RL240189S, RL240146A, RL240154A, RL240173A, RL240181A, RL240203A, RL240194A, RL240200A, RL240212A, RL240167A, RL240213A, RL240206A, RL240221A, RL240214A, RL250007A, RL240136A, RL250008A, RL240233A,...
- Reason
- In a specific software version of this system, a bug was found that the manually corrected "coronary artery height" value in the heart valve analysis option was reset to the automatically extracted value after a specific operation. Therefore, we have decided to correct the software as a voluntary modification.
- Classification
- Class II
- Category
- medical device recall
- Country
- Japan
- Agency
- PMDA-DEVICE
- Recall date
- 2026-02-20
- Status
- Ongoing
Official recall notice from PMDA-DEVICE