Medline Industries, Lp: Catheter Care Tray
Medline Catheter Care Tray Recall Summary Medline Industries is recalling Pre-connected Urethral Catheterization Trays (15 Fr, red rubber latex) because some products have surface irregularities on the catheter, connectors that may detach from the catheter funnel, or catheters that don't meet size specifications. The recall affects products distributed nationwide in the US and several other countries.
- Recalling firm
- Medline Industries, LP
- Product
- MEDLINE Pre-connected Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND10407
- Reason
- During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-05-04
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/ff24c288-ca01-4f39-892b-71c28bb8a488