Philips North America: Philips Achieva 1.5t Initial System With Mr Elastography (mre). 1. Model Number (ref): 781178

Recall Summary Philips is recalling certain Achieva 1.5T MRI machines (Model 781178) because specific combinations of image reconstruction settings can produce inaccurate stiffness measurements when using MR Elastography technology. This affects units distributed worldwide, including throughout the United States and multiple other countries. This is classified as a Class II recall, meaning the problem could result in incorrect diagnostic information but is not an immediate safety threat.

Recalling firm
Philips North America
Product
Philips Achieva 1.5T Initial system with MR Elastography (MRE). 1. Model Number (REF): 781178.
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-04-14
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar,...

Official recall notice from U.S. Food and Drug Administration

Recall details on Nibble Recalls: https://trynibble.app/recalls/fdb4d0d4-b930-4fad-a379-a6ed7021aa8f