Asp Global, Llc.: Tourniquet, Pneumatic
Recall Summary ASP Global is recalling MPR brand single-use tourniquet cuffs (model DT-7108-02Q) manufactured between September 2016 and March 2023 because they are labeled with a 10-year shelf life when the actual validated shelf life is only 5 years. These sterile devices are used during surgery to reduce blood flow to a patient's limb, and the incorrect expiration dates mean some units may no longer be sterile or effective. The affected products were distributed nationwide across most U.S. states.
- Recalling firm
- ASP Global, LLC.
- Product
- Brand Name: MPR Product Name:SINGLE USE TOURNIQUET CUFF 8" 2 LINE (10) Model/Catalog Number: DT-7108-02Q Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No
- Reason
- Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
- Classification
- Class I
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-08-24
- Status
- Open, Classified