Philips North America: Philips Ingenia 3.0t With Mr Elastography (mre). 1. Model Number (ref): 781342. 2. Model Number (ref): 781377. 3. Model Number (ref): 782103. 510(k...

Recall Summary Philips is recalling three models of its Ingenia 3.0T MR Elastography machines (model numbers 781342, 781377, and 782103) distributed worldwide because they may produce incorrect stiffness measurements when certain image settings are used together. The error occurs due to how the reconstruction voxel size is configured in the default scan protocol when using Resoundant's algorithm.

Recalling firm
Philips North America
Product
Philips Ingenia 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781342. 2. Model Number (REF): 781377. 3. Model Number (REF): 782103. 510(k) Numbers: K163116, K173079, K183063, K193215, K213516, K213583, K223458, K251397, K251808
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-04-14
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar,...