Retractor Components
- Recalling firm
- Integra Neurosciences Pty Ltd (NZ)
- Product
- Retractor Components. [Device]. Model: Integra Omni-Tract Surgical Retractor System; Integra Omni-Tract Surgical Retractor System; Integra Omni-Tract Surgical Retractor System. Affected: All lot numbers; All lot numbers; All lot numbers. Also covers: Omni-Tract Radiolucent; Omni-Tract Obesity Blades
- Reason
- An Instructions for Use (IFU) update has been issued for the Omni-Tract Surgical Retractor System to update validated steam sterilisation parameters and approved configurations. The previous IFU specified a 30-minute drying time; however, validation has confirmed that longer drying times (90–120 minutes depending on cycle exposure) are required to ensure adequate drying in pre-vacuum steam sterilisation cycles., The correction also removes rigid container use and gravity displacement sterilisation as approved methods, as these configurations have not been validated., This update is to...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-06-29
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE