Cardioroot

Recalling firm
Getinge Australia Pty Ltd (NZ)
Product
Cardioroot. [Device]. Model: CARDIOROOT Woven - Model number: HEWROOT0024, HEWROOT0026, HEWROOT0028, HEWROOT0030, HEWROOT0032, HEWROOT0034. Affected: All CARDIOROOT Woven products within labelled shelf life
Reason
Internal testing of samples from CARDIOROOT vascular grafts produced Integral Water Permeability (IWP) results outside the limit specified in the product's Instructions for Use (IFU) (i.e. <5 ml/cm²/min). Given that the affected products did not meet the specified acceptance criteria for water permeability, they are out of specification and may pose risks to patient health.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-18
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/fb396cc2-2b48-484a-984e-97480f240ad0