Instrumentation Laboratory: Electrode Measurement, Blood-gases (pco2, Po2) And Blood Ph

Recall Summary Instrumentation Laboratory is recalling GEM Premier 7000 blood gas analyzers with iQM3 quality control systems that received a specific software update on or after April 8, 2026. A software error can cause patient test results to be assigned to the wrong patient ID, meaning blood work from one patient could be labeled with another patient's information. This recall affects devices distributed nationwide across all U.S. states and the District of Columbia.

Recalling firm
Instrumentation Laboratory
Product
Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000.
Reason
BCR wands that received a configuration update (released under Service Information No. 2026-07 on or after April 8, 2026) may assign patient results to an incorrect patient ID. The root cause is a race condition introduced when the "price check" feature was disabled to reduce excess current draw. This causes buffered data from a previous scan to be transmitted instead of the most recent scan, populating a prior patient ID instead of the current one. This can result in erroneous results being attributed to the wrong patient, potentially leading to inappropriate clinical management.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-08
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/fb2ebd5e-a434-4797-b2e7-e6dde461b4a7