Aniara Diagnostica Llc: Plasma, Coagulation Control

Recall Summary Aniara Diagnostica LLC is recalling its 5-PLASMA Bivalirudin Control (Model 5D-44905), a medical device used to test blood coagulation, because the reconstituted product breaks down over time and loses measurable bivalirudin levels, which can affect test accuracy. The product was distributed nationwide, primarily in Illinois, North Carolina, and Tennessee. This is a Class II recall, indicating a potential issue that could affect proper device function.

Recalling firm
Aniara Diagnostica LLC
Product
5-PLASMA Bivalirudin Control, Model Number 5D-44905
Reason
Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-07-01
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of IL, NC, TN.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/fa5843c3-7f8f-46e5-9148-7b1a45c3431c