[mdr]philips Patient Information Centre Ix Pic Ix
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- [MDR]Philips Patient Information Centre iX PIC iX. [Device]. Model: Philips Patient Information Centre iX (PIC iX) 4.x systems
- Reason
- The manufacturer received reports that some PIC iX systems were frozen and could not be used. It was determined that failure of the watchdog may cause the system to be in an unknown state and therefore potentially prevent alarms from being generated or logged.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-08-24
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/f988570a-5670-422b-a246-29666284b68f