[mdr]philips Patient Information Centre Ix Pic Ix

Recalling firm
Philips New Zealand Commercial Limited
Product
[MDR]Philips Patient Information Centre iX PIC iX. [Device]. Model: Philips Patient Information Centre iX (PIC iX) 4.x systems
Reason
The manufacturer received reports that some PIC iX systems were frozen and could not be used. It was determined that failure of the watchdog may cause the system to be in an unknown state and therefore potentially prevent alarms from being generated or logged.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-24
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory