Erbe USA Inc: Unit, Cryosurgical, Accessories

Recall Summary Erbe USA Inc is recalling the ERBECRYO 2 Cryosurgical Unit (Model 10402-000) because pulling forcefully on the high-pressure transfer hose can cause a gas leak that builds up excessive pressure inside the device, potentially rupturing the cryoprobe. The affected units were distributed nationwide across all U.S. states and territories.

Recalling firm
Erbe USA Inc
Product
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
Reason
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-26
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts,...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/f8801af1-5f22-4220-8a47-dfbb278fb931