I.v. Start Kit
Recall Summary Medline is recalling six medical kits (including IV catheter trays, arthroscopy kits, and feeding tube kits) due to calibration problems with sterilization equipment that may have compromised the sterility of the products. The affected kits were distributed nationwide across all U.S. states and territories. Healthcare facilities and patients who received these products should check if they have any of the recalled items, which are identified by specific model numbers.
- Product
- Medline Convenience Kits: 1) PICC CATHETER INSERTION TRAY, Model Number: CVI4300; 2) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977C; 3) KIT ACES INSERT PEG, Model Number: DYKMBNDL130; 4) KIT ACES INSERT PEG, Model Number: DYKMBNDL130A; 5) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4; 6) NICU LINE CHANGE TRAY, Model Number: LC295
- Reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-01-07
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide distribution.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/f8388920-4f6b-4b23-a54a-9b2176bfb689