Alphenix (all Models)
Alphenix Recall Summary Canon Medical Systems Corporation is recalling all Alphenix models distributed to hospitals, pharmacies, and laboratories in New Zealand due to a need for a software upgrade. This is classified as a Class I recall, which indicates a serious issue. The recall affects the Alphenix equipment in medical facilities across New Zealand.
- Recalling firm
- Canon Medical Systems ANZ Pty Limited
- Product
- Alphenix (all models). [Device]. Model: Angiography system (INFX-8000C; INFX-8000V)
- Reason
- The manufacturer has identified that an internal error may occur when CAAS Clinical Functions are used during an examination, potentially causing a system lockup and loss of essential functions such as fluoroscopy and digital acquisition. Although a partial reset is recommended, the system may fail to restart due to temporary memory or storage read errors, leading to procedure interruptions or treatment delays that require a system restart.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-05-01
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory