Alphenix (all Models)

Alphenix Recall Summary Canon Medical Systems Corporation is recalling all Alphenix models distributed to hospitals, pharmacies, and laboratories in New Zealand due to a need for a software upgrade. This is classified as a Class I recall, which indicates a serious issue. The recall affects the Alphenix equipment in medical facilities across New Zealand.

Recalling firm
Canon Medical Systems ANZ Pty Limited
Product
Alphenix (all models). [Device]. Model: Angiography system (INFX-8000C; INFX-8000V)
Reason
The manufacturer has identified that an internal error may occur when CAAS Clinical Functions are used during an examination, potentially causing a system lockup and loss of essential functions such as fluoroscopy and digital acquisition. Although a partial reset is recommended, the system may fail to restart due to temporary memory or storage read errors, leading to procedure interruptions or treatment delays that require a system restart.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-05-01
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE