Lp(a) Calibrator Series

Recalling firm
Banksia Scientific Pty Ltd (NZ)
Product
Lp(a) Calibrator Series. [Device]. Model: Catalogue number: LP3404. Affected: Batch number: 899841 Lot: 1955LP, 1956LP, 1957LP, 1958LP, 1959LP
Reason
The manufacturer is recalling LP(a) Calibrator Series, catalogue number LP3404, batch 899841, lots 1955LP, 1956LP, 1957LP, 1958LP and 1959LP, due to a quality issue affecting a limited number of vials. A limited number of vials are cloudy in appearance and the level of Lp(a) in these vials is impacted. If a calibration is performed using one of the impacted calibrators the calibration curve will be impacted. In some cases, calibration may fail, but calibrations may also pass and patient recovery may be impacted., Due to the variation between vials and subsequent impact on calibration and...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-17
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE