Lp(a) Calibrator Series
- Recalling firm
- Banksia Scientific Pty Ltd (NZ)
- Product
- Lp(a) Calibrator Series. [Device]. Model: Catalogue number: LP3404. Affected: Batch number: 899841 Lot: 1955LP, 1956LP, 1957LP, 1958LP, 1959LP
- Reason
- The manufacturer is recalling LP(a) Calibrator Series, catalogue number LP3404, batch 899841, lots 1955LP, 1956LP, 1957LP, 1958LP and 1959LP, due to a quality issue affecting a limited number of vials. A limited number of vials are cloudy in appearance and the level of Lp(a) in these vials is impacted. If a calibration is performed using one of the impacted calibrators the calibration curve will be impacted. In some cases, calibration may fail, but calibrations may also pass and patient recovery may be impacted., Due to the variation between vials and subsequent impact on calibration and...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-11-17
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/f5d95e01-6f30-41d4-9365-f4f03b27b3c9