Bvi Leos Laser And Endoscopy System. Model Number: Ome6000u.

Recall Summary The BVI Leos Laser and Endoscopy System (Model OME6000U) is being recalled because the USB cable connecting the Nurse Monitor touchscreen to the System Controller is missing ferrite components that were supposed to be installed at both ends of the cable. This manufacturing error means the device doesn't meet its design specifications. The recall affects all users of this system model.

Recalling firm
Beaver-Visitec International, Inc.
Product
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Reason
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
Classification
Class II
Category
consumer product recall
Country
United States
Agency
FDA-DEVICES
Recall date
2025-12-05
Status
Open, Classified

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/f5beface-b11d-4657-9141-7f6946b891a9