[mdr] Spectral Ct, Iqon Spectral Ct, Brilliance Ict, Brilliannce Ict Upgrade

Recalling firm
Philips New Zealand Commercial Limited
Product
[MDR] Spectral CT, IQon Spectral CT, Brilliance iCT, Brilliannce iCT Upgrade. [Device]. Model: Spectral CT
Reason
As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-30
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory