Generic Name: (1) Digital Extraoral General-purpose Dental X-ray Diagnostic System (analog Extraoral General-purpose Dental X-ray Diagnostic System) (2) Analog — Extraoral General-purpose Dental X-ray Diagnostic System — Trade Name: (1) Port X IV — (2) Port X Ive

Recalling firm
Name of Manufacturer and Distributor: Morita Co., Ltd. — Address of Manufacturer and Distributor: 3-33-18 Tarumi-cho, Suita City, Osaka Prefecture — Type of License: Class I Medical Device...
Product
Lot: 販売名 :(1)ポートエックスIV 対象ロット: GD-526016-20625,GD-526020-20625,GD-526022-20625,GD-526004-20625 GD-526001-20625,GD-507001-20625,GD-526002-20625,GD-526007-20625 GD-526018-20625,GD-526030-20625,GD-526028-20625,GD-526003-20625 GD-526021-20625,GD-526017-20625,GD-526019-20625,GD-526029-20625 GD-526032-20625,GD-526011-20625,GD-526012-20625,GD-526013-20625 GD-526015-20625,GD-526031-20625 数量及び出荷時期: 22台 令和7年7月14日 〜 令和7年9月8日 販売名 :(2)ポートエックスIVe 対象ロット: GD-602023-20625,GD-602024-20625,GD-328001-20625,GD-328003-20625 GD-328008-20625,GD-328002-20625,GD-407002-20625,GD-602026-20625...
Reason
We have received reports from the market that batteries are unable to charge. Following an investigation by the manufacturer, it was determined that in this particular lot, if the power source is connected after the battery has naturally discharged, a short circuit may occur within the charging circuit board, preventing the battery from charging. Furthermore, since the short circuit may cause minor sparks and smoke, we have decided to conduct a recall.
Classification
Class II
Category
medical device recall
Country
Japan
Agency
PMDA-DEVICE
Status
Ongoing

Official recall notice from PMDA-DEVICE