Canon Medical System, Usa, Inc.: Interventional Fluoroscopic X-ray System

Recall Summary Canon Medical System's ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM (model INFX-8000F) X-ray equipment is being recalled due to a software issue that can cause internal errors when certain functions are used simultaneously during medical procedures. The affected systems were distributed nationwide to healthcare facilities. This is classified as a Class II recall, meaning the issue could potentially impact device performance but does not present an immediate serious health risk.

Recalling firm
Canon Medical System, USA, INC.
Product
ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000F
Reason
Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-10-12
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/f03aaaea-0fef-4095-b5db-c1d38c0d5850