Boston Scientific Neuromodulation Corporation: Stimulator, Spinal-cord, Implanted (pain Relief)

Recalling firm
Boston Scientific Neuromodulation Corporation
Product
Nalu TD2 Therapy Disc with firmware, REF: 34014-001, and TD2 Patient Kit, REF: 74011, part of the Nalu Neurostimulation System
Reason
Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-09-30
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.