Datex-ohmeda Inc: Warmer, Infant Radiant

Recall Summary Datex-Ohmeda is recalling certain Panda Freestanding Infant Radiant Warmers that have a specific blender component installed during recent maintenance. The blender knob shaft on this component can loosen, causing the warmer to deliver oxygen at a concentration different from what the operator set, which could result in too little or too much oxygen being delivered to the patient. This recall affects only units with the specific service blender (M1091607-R) that was installed during the last preventative maintenance.

Recalling firm
Datex-Ohmeda Inc
Product
Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.
Reason
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-08
Status
Open, Classified
Distribution
Pending

Official recall notice from FDA-DEVICES