Aniara Diagnostica Llc: Calibrator, Multi-analyte Mixture

Recall Summary Aniara Diagnostica LLC is recalling its 5-PLASMA Bivalirudin Calibrator (Model 5D-46905), a product used to calibrate blood testing equipment. The calibrator's reconstituted plasma mixture degrades over time, causing the bivalirudin measurement levels to decrease and potentially produce inaccurate test results. The product was distributed nationwide, with primary distribution in Illinois, North Carolina, and Tennessee.

Recalling firm
Aniara Diagnostica LLC
Product
5-PLASMA Bivalirudin Calibrator, Model Number 5D-46905
Reason
Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-07-01
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of IL, NC, TN.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/ecd62d59-9fea-48af-82ee-2994952b5d8f