Clearview Hcg Cassette And Innovacon Hcg One Step Pregnancy Test Device

Recalling firm
Abbott Laboratories (NZ) Ltd
Product
Clearview hCG Cassette and Innovacon hCG One Step Pregnancy Test Device. [Device]. Model: Clearview™ hCG Cassette (20 mIU/mL), Innovacon™ hCG One Step Pregnancy Test Device (Urine). Affected: Part number: 92210, FHC-102 - All lots within expiration date
Reason
Update 2/9/26: In consideration of the current market shortage of the affected products, this action has been revised for distributors. Distributors may continue to supply inventory to their customers, on the provision that all future customer orders must be accompanied by the Urgent Product Correction Customer Letter and Customer Reply Form., As part of routine post-market surveillance activities, an investigation was conducted into reports of false-positive results, and determined that while the tests demonstrate expected results at and above the established sensitivity levels, they may...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-09-02
Status
Ongoing
Distribution
New Zealand - Healthcare Professional

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/e9b62487-2767-446c-8b56-54a7c1939399