Arrow International, Llc: Catheter, Hemodialysis, Implanted
Recall Summary Arrow International is recalling specific Cannon II Plus Hemodialysis Catheter kits because a component called a sheath introducer may not split apart properly during use. This defect could cause difficulty removing the catheter, longer procedure times, and potentially lead to pain, bleeding, bruising, or tissue damage. The affected products were distributed nationwide in the US and to numerous countries worldwide.
- Recalling firm
- ARROW INTERNATIONAL, LLC
- Product
- Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15242-SPM, CS-15282-SPM, CS-15322-SPM, CS-15362-SPM
- Reason
- 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-04-10
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution: US nationwide and countries of: AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR,...
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/e98fc0d7-4323-446e-97d6-67529f5afb59