B Braun Medical Inc: B Braun Hemodialysis Bloodlines, Low Vol, No Chamber 4.8mm Pump Segment. Model Number: B3-4630m4705
- Recalling firm
- B Braun Medical Inc
- Product
- B Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 4.8MM PUMP SEGMENT. Model Number: B3-4630M4705.
- Reason
- Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.
- Classification
- Class I
- Category
- medical device recall
- Country
- United States
- Agency
- U.S. Food and Drug Administration
- Recall date
- 2026-03-19
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
Official recall notice from U.S. Food and Drug Administration
Recall details on Nibble Recalls: https://trynibble.app/recalls/e97d3f67-4cfa-4f42-b00b-baafecb028f1