Circuit, Breathing (w Connector, Adaptor, Y Piece)

Recall Summary Medline is recalling breathing circuit kits (two models: DYNJ0185290G and DYNJ905567A) used in anesthesia because equipment problems during sterilization may have compromised the sterility of the products. The affected kits were distributed nationwide across all U.S. states and territories.

Product
Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number:DYNJ0185290G; 2) OPEN HEART ANESTHESIA ICC-A, Model Number:DYNJ905567A
Reason
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-01-07
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution.

Official recall notice from FDA-DEVICES