Abiomed, Inc.: Introducer, Catheter
Recall Summary Abiomed is recalling Introducer Kits (medical devices used with heart pumps) in the US and worldwide due to potential leakage from the introducer sheath that cannot be seen by the user. The issue affects 23 Fr and 14 Fr introducers, which are included in every pump set sold. This is a Class II recall, meaning the defect could cause serious health problems but is unlikely to result in death.
- Recalling firm
- Abiomed, Inc.
- Product
- ABIOMED Introducer Kit 23 Fr, 11 cm, Pump Set Code 1000441, containing Kit, 23Fr Introducer, 11cm, Sterile, Introducer Product Code 2000342. heart pump and accessories
- Reason
- A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-06-11
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT),...