Baxter Healthcare Corporation: Device, Positive Pressure Breathing, Intermittent

Recall Summary Baxter Healthcare is recalling the Hillrom VOLARA System P.CIRCUIT KIT, a positive pressure breathing device used for lung expansion therapy, because the nebulizer cup may not lock properly after medication is added, causing air and medication to leak during treatment. This leakage can reduce oxygen levels and prevent proper medication delivery to patients. The recall affects devices distributed nationwide in the US and Canada.

Recalling firm
Baxter Healthcare Corporation
Product
Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device
Reason
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-21
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the country of Canada.