Apexpro Carescape Telemetry Server

Recalling firm
GE Healthcare Limited
Product
ApexPro CARESCAPE Telemetry Server. [Device]. Model: ApexPro CARESCAPE Telemetry Server v5.0 and v6.0 with CARESCAPE Central Station or CIC Pro Clinical Information Center
Reason
The manufacturer has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-05-15
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/e68c7519-0839-48c3-9d46-05767f739574