Philips North America: System, Nuclear Magnetic Resonance Imaging

Recall Summary Philips is recalling Achieva 3.0T MRI machines with MR Elastography (model 781278) distributed worldwide because certain image reconstruction settings can produce inaccurate stiffness measurements, caused by the voxel size settings in the default scan protocol displaying incorrectly. This is a Class II recall affecting machines across all U.S. states and numerous countries including Canada, the United Kingdom, and many others.

Recalling firm
Philips North America
Product
Philips Achieva 3.0T with MR Elastography (MRE). 1. Model Number (REF): 781278.
Reason
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-14
Status
Open, Classified
Areas affected