Ge Medical Systems, Llc: System, X-ray, Mobile

Recall Summary GE HealthCare is recalling certain Optima XR200/220/240amx mobile X-ray systems that have undergone column replacement, because safety testing may not have been completed after the repair. If an internal cable fails, the arm holding the X-ray tube could drop unexpectedly, potentially injuring patients during imaging procedures. The recall affects systems distributed nationwide and in several other countries.

Recalling firm
GE Medical Systems, LLC
Product
Optima XR220amx, Mobile Digital-Ready Radiographic System
Reason
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-05
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/e203ac59-3224-45aa-99c9-185542d92c0f