Smiths Medical Asd (mmsp) Hotline Fluid Warming Set

Recalling firm
Smiths Medical Australasia Pty Ltd
Product
Smiths Medical ASD (MMSP) HOTLINE Fluid Warming Set. [Device]. Model: Hotline Fluid Warming Set. Affected: Lot number - 4293569, 4337725
Reason
The manufacturer became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond. Two (2) Lots of L-70 HOTLINE Fluid Warming Sets have an increased likelihood of leakage due to this problem.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-07-02
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/e0b33cd4-13e3-4632-98ba-7ddb6a91ee5e