Nobel Biocare Implant Extractor

Nobel Biocare AB is recalling the Nobel Biocare Implant Extractor device in New Zealand. The device is being returned to the supplier. This recall affects healthcare professionals who use this product.

Recalling firm
Nobel Biocare New Zealand Ltd
Product
Nobel Biocare Implant Extractor. [Device]. Model: Implant Retrieval Instrument CC RP & Tri-Channel WP 22mm. Affected: Lot number - 76719
Reason
Due to a process error at the manufacturer, the Implant Retrieval Instrument CC RP & Tri-Channel WP 22mm has been mixed up with an Implant Retrieval instrument intended for the wide platform (WP). The laser marking on the device (CC RP) is in line with the packaging, but not correct for the delivered article., The affected Implant Retrieval Instrument does not function as intended. Since the connecting part of the instrument is too wide, it cannot be used to retrieve a CC RP or Tri-Channel WP implant.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-05-08
Status
Ongoing
Distribution
New Zealand - Healthcare Professional

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/e08d0623-ae17-4453-9f61-7d1a63cb0592