Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(ous Only); Software Version Number - All Versions: R1.x, R2.x, R3.x

Recall Summary Philips is recalling certain Azurion 5 M20 imaging systems (models 722228, 722232, and 722281) with software versions R1.X, R2.X, and R3.X because the examination table may move unexpectedly when the Reset Geometry button is pressed, even if the table lock is engaged. This is classified as a Class II recall, which indicates a serious potential health risk. The recall affects medical facilities using these specific system models and software versions.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
Product
Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Reason
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
Classification
Class II
Category
consumer product recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-01-26
Status
Open, Classified

Official recall notice from FDA-DEVICES