Avalon Fetal Monitors
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- Avalon Fetal Monitors. [Device]. Model: Philips Avalon FM20 Fetal Monitor M2702A, Avalon FM30 Fetal Monitor M2703A
- Reason
- The manufacturer has identified that the Avalon Speaker Assembly cable contacts were inserted, rotated 180° relative to the orientation specified. The reversed contacts in the Connector-Housing cause the locking nose to fail to engage the Connector Socket, which permits the cables to be pulled out of the housing and increases the risk of intermittent or permanent loss of speaker output. If users do not routinely have fetal heart rate tones enabled, it may take longer for a speaker malfunction to be identified prior to the occurrence of harm. If audible fetal heart rate tones are enabled,...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-06-18
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE