D Dimer

Recalling firm
Diagnostica Stago Pty Ltd
Product
D Dimer. [Device]
Reason
After receiving customer complaints, the manufacturer conducted thorough investigations and confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range., This bias affects the specificity values reported in the test's package insert. The recalculated specificity is 63.9% for Pulmonary Embolism (compared to 75.5% stated in the package insert) and 41.2% for Deep Vein Thrombosis (compared to 55.2% stated in the package insert).
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-02-12
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory