Efficia Dfm100, Heartstart Intrepid
Recall Summary Philips is recalling Efficia DFM100 Defibrillator/Monitor devices with software version 2.00.33 because they may fail to start up properly and become unusable during critical moments. The affected devices will repeatedly show a splash screen instead of fully booting, indicated by a red X in the Ready For Use indicator. This recall affects devices distributed to hospitals, pharmacies, and laboratories in New Zealand.
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- Efficia DFM100, HeartStart Intrepid. [Device]. Model: Efficia DFM100 devices with software revision 2.00.33 installed. Affected: Boot-Up Failure with SW 2.00.33
- Reason
- The manufacturer has identified a potential safety issue with the Efficia DFM100 Defibrillator/Monitor running software version 2.00.33. Affected devices may fail to complete their boot-up sequence. This can occur either:, during manual boot-up when the knob is turned to AED, Manual, or Pacer mode, or, during a self-test when the device is powered off and connected to any power source., Instead of fully booting, the device may repeatedly display the splash screen and will not become available for clinical use. This condition is indicated to the user by a solid red X shown in the Ready For...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-03-10
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/d8a246a7-bf56-45fc-a086-e2bbf0bb5e7a