Angiographic X-ray
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- Angiographic X-Ray. [Device]. Model: a Philips Azurion R2.x.10, R3.0, R3.1 and R4.0 systems; Azurion R3.0 systems are affected by these issues. Issue 2 - Loss of geometry movements only affects Azurion R3.0 systems equipped with an AD7X patient table.. Also covers: [MDR] Azurion
- Reason
- The manufacturer has identified 2 potential problems that may lead to delay of therapy, procedural complications and/or inappropriate treatment., · Problem 1: Image quality degradation, A problem associated with the Flat Detector Controller System Interface Board (FDCSIB), may result in image artifacts during, system operation. The observed artifacts may include black and white dot noise, grainy or cloudy image quality, snow-like, effects, and random pixelation., · Problem 2: Loss of geometry movements – Applicable to systems with an AD7X table only, In some cases, following a system...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-08-10
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE