Recalling firm
Philips New Zealand Commercial Limited
Product
Angiographic X-Ray. [Device]. Model: a Philips Azurion R2.x.10, R3.0, R3.1 and R4.0 systems; Azurion R3.0 systems are affected by these issues. Issue 2 - Loss of geometry movements only affects Azurion R3.0 systems equipped with an AD7X patient table.. Also covers: [MDR] Azurion
Reason
The manufacturer has identified 2 potential problems that may lead to delay of therapy, procedural complications and/or inappropriate treatment., · Problem 1: Image quality degradation, A problem associated with the Flat Detector Controller System Interface Board (FDCSIB), may result in image artifacts during, system operation. The observed artifacts may include black and white dot noise, grainy or cloudy image quality, snow-like, effects, and random pixelation., · Problem 2: Loss of geometry movements – Applicable to systems with an AD7X table only, In some cases, following a system...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-10
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE