Gentuity, Llc: Gentuity Hf-oct Imaging System Model G10-01; Software Version 23.2.13

Gentuity HF-OCT Imaging System Recall Summary Gentuity is recalling the HF-OCT Imaging System Model G10-01 with software version 23.2.13 because the system may repeat frames during imaging pullback when used with certain catheters, which could affect the accuracy of length measurements. This issue affects medical facilities in the US (across most states), Belgium, and Japan that have this specific equipment and software version. The company is providing information to help users recognize and prevent this condition during use.

Recalling firm
Gentuity, LLC
Product
Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
Reason
The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-03-23
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Domestic - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN, NJ, NY, OH, PA, TX; International: United Arab Emirates, Belgium, Switzerland, Cyprus, Denmark, Spain, Finland, France, Greece, Italy, Japan, Kuwait, Latvia, Poland, Saudi Arabia, Slovakia;

Official recall notice from U.S. Food and Drug Administration

Recall details on Nibble Recalls: https://trynibble.app/recalls/d6eb84be-6425-4ef0-9554-48d2841dbdeb