Celon Ent Electrodes

Recalling firm
Olympus (NZ) Ltd
Product
Celon ENT Electrodes. [Device]. Model: CelonProBreath - Model - WB990004, WB990007. Affected: Multiple lot numbers
Reason
The manufacturer has identified that certain boxes of CelonProBreath bipolar applicator was released to customers without the required Instructions for Use (IFU) included. Investigation concluded that the cause of the issue was a software configuration error in the packaging process which allowed these devices to be packaged without the IFU. The software configuration error has since been resolved.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-02-09
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/d5e5c70c-2dbb-4891-a467-1e9fde780ede