Philips Azurion Systems

Recalling firm
Philips New Zealand Commercial Limited
Product
Philips Azurion systems. [Device]. Model: Module name: Azurion 7 M20. Affected: Product number: 722079
Reason
The manufacturer has identified that, in certain Azurion systems, one or more of these power supplies can fail due to energy peaks (surge voltages) during start or restart of the system. When a power supply failure occurs during a start or restart, the Power Distribution Unit does not provide the required direct current power to essential system components. As a result, the system does not successfully complete the (re-)start, imaging functionality is unavailable, and the system remains unavailable for clinical use.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-04
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE