Icu Medical Inc.: Hotline Fluid Warming Set, Single Use Device, Is Designed To Warm Blood And Intravenous Fluids And Deliver Them To The Patient's Intravenous Access...

Recall Summary ICU Medical is recalling certain lots of its Hotline Fluid Warming Set, a single-use device that warms blood and IV fluids before delivering them to patients. The sets may leak due to a manufacturing defect where solvent was not applied adequately at the connection between the tubing and return connector. The recall affects products distributed nationwide across all U.S. states and territories.

Recalling firm
ICU Medical Inc.
Product
Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70
Reason
ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-06-24
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide - US Nationwide distribution.

Official recall notice from U.S. Food and Drug Administration

Recall details on Nibble Recalls: https://trynibble.app/recalls/d27f4bc0-b827-4bad-8bb3-01569fc760b6