一般的名称: 体動センサ 販売名 : 眠りscan(nn‐1530)

Recall Summary Paramount Bed Co., Ltd. is recalling 47 units of the Nemuri SCAN (NN-1530), a body motion sensor that measures sleep by detecting movement, breathing, and heart rate. The devices were shipped between January 20 and February 25, 2026, across specific manufacturing lot numbers. The sensor may display an incorrect MAC address on its label due to a firmware error, which can cause the device to malfunction and fail to take measurements. No health injuries have been reported, and the manufacturer states that measurement failure poses no serious health risk.

Recalling firm
製造販売業者の名称 : パラマウントベッド株式会社 千葉工場 製造販売業者の所在地: 千葉県山武市白幡2078 許可の種類 : 第一種医療機器製造販売業 許可番号 : 12B1X10020
Product
Lot: 製造番号: 021386600046,021387200137,021387200138,021387200139,021387200140,021387200142, 021387200146,021387200149,021387200151,021387200216,021387200217,021387200219, 021387200220,021387200221,021387200225,021387200227,021387200229,021387300022, 021387300041,021387300042,021387300044,021387300046,021387300047,021387300056, 021387300068,021387400201,021387700064,021388600174,021388600178,021388600184, 021388600189,021388600211,021388600212,021388600216,021388600218,021388600219, 021389000063,021389200209,021389200253,021389300075,021389300078,021389300093,...
Reason
当該機器を設置する際、エラーが発生して正常に動作しない事象が確認されました。調査の結果、本体に貼付し たラベルに誤ったMACアドレスが表示されている場合のあることが判明しました。MACアドレスに特定の文 字列が含まれている場合に、ファームウェアの不具合により、MACアドレスを正しく読み取れず、誤った情報 をもとにラベルが印刷される事象が発生していることが判明したため、対象となる製品について、回収すること としました。
Classification
Class II
Category
medical device recall
Country
Japan
Agency
PMDA-DEVICE
Recall date
2026-07-06
Status
Ongoing

Official recall notice from PMDA-DEVICE