Senographe Pristina Via Systems

Recalling firm
GE Healthcare Limited
Product
Senographe Pristina VIA systems. [Device]. Model: Senographe Pristina VIA 3D+2D. Affected: Certain clauses of IEC60601-2-45 Edition 3.1/3.2 standard
Reason
The manufacturer has become aware of a nonconformance to certain clauses of IEC60601-2-45 Edition 3.1/3.2 standard in some Senographe Pristina VIA systems., Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following:, � the collimator blades might be visible in the image, or the collimator might be open more than expected, and, � an incorrect collimator filter might be used with respect to the anode track selected, This issue could result in a minor additional X-ray dose being delivered...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-10-13
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE