Palamix Cement Dispenser

Palamix Cement Dispenser Recall Summary Palamix Cement Dispensers (uno and duo models) with batch numbers below 223060 and 520215 respectively are being recalled because their printed expiration dates are incorrect. The shelf life for these earlier batches should be 18 months shorter than what is printed on the label. The affected products were distributed to hospitals, pharmacies, and laboratories in New Zealand.

Recalling firm
Orthomed NZ Ltd
Product
Palamix Cement Dispenser. [Device]. Affected: batch number - 223060 (PALAMIX® uno) and 520215 (PALAMIX® duo)
Reason
Beginning with batch number 223060 (PALAMIX® uno) and 520215 (PALAMIX® duo), the new shelf life of 1.5 years will be shown on the product label. Since batch numbers are assigned consecutively, all batches numbered below 223060 (PALAMIX® uno) and below 520215 (PALAMIX® duo) will continue to display the original shelf life. For these earlier batches, the printed expiration date must be reduced by 18 months.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-02-19
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory