Generic Name: (1)(2) Positron Ct Hybrid Spect System, (3) Rotating Detector Spect System For Nuclear Medicine Diagnostics — Trade Name: (1) Millennium Vg — (2) Infinia — (3) Infinia Hawkeye 4

Recall Summary GE Healthcare is recalling three nuclear medicine imaging systems (Millennium VG, Infinia, and Infinia Hawkeye 4) used in hospitals and diagnostic centers due to potential mechanical failures that could cause the detector to descend unintentionally. The affected equipment was manufactured between 2000 and 2016, and while the devices have alarm systems to alert staff if malfunctions occur, no patient injuries have been reported to date.

Recalling firm
Name of Manufacturer/Distributor: GE Healthcare Japan Co., Ltd. — Address of Manufacturer/Distributor: 4-7-127 Asahigaoka, Hino City, Tokyo — License Type: Class I Medical Device...
Product
Lot: (1)ミレニアムVG 製造番号 : YN0498、YN0984 数量 : 2台 出荷時期 : 平成12年 2月〜平成15年 2月 (2)インフィニア 製造番号 : YN1196、YN1540、YN1544、YN1545、YN1546、YN1547、YN1548、YN1551、YN1552、YN1554、 YN1555、YN1556、YN1558、YN1559、YN1561、YN1562、YN1566、YN1567、YN1572、YN1573、 YN1575、YN1578、YN1579、YN1580、YN1581、YN1582、YN1583、YN1584、YN1585、YN1586、 YN1587、YN1588、YN1589、YN1590、YN1592、YN1601、YN1641、YN1744、YN4219、YN4227、 YN4230、YN4231、YN4233、YN4237、YN4240、YN4243、YN4244、YN4248、YN4249、YN4251、 YN4252、YN4253、YN4255、YN4256、YN4257、YN4259、YN4260、YN4261、YN4262、YN4263、 YN4264、YN4265、YN4266、YN4267、YN4280、YN4288、YN4312、YN4818、YN4819、YN4820、...
Reason
We have been informed by the manufacturer that if mechanical failures—such as malfunctioning radial gears, gear oil leaks, or misaligned brake pads—occur on this equipment while the correct version of the gantry software is not installed, the detector may descend slowly, potentially causing unintended operation. Therefore, based on information from the manufacturer, we will provide information to our customers, inspect the gantry software version and brake components, and perform corrective work to address any issues found.
Classification
Class II
Category
medical device recall
Country
Japan
Agency
PMDA-DEVICE
Status
Ongoing

Official recall notice from PMDA-DEVICE

Recall details on Nibble Recalls: https://trynibble.app/recalls/cfaa3851-30b3-4b6c-83bd-447ac8e6df9d