Airseal® Ifs Access Port With Trocar

Recalling firm
Paragon Care Group New Zealand Limited
Product
AirSeal® iFS Access Port with Trocar. [Device]. Affected: Catalogue number - iASB5-150, iASB12-120, iASB12-100, iAS8-120LP, iAS8-100LP, iAS5-75LP, iAS5-120LP, iAS5-100LP, iAS12-150LPI, iAS12-120LPI, iAS12-100LPI
Reason
The manufacturer has received an increase in reports of overpressure alerts with AirSeal Access Ports, which in some cases have resulted in delay of treatment. During a review of complaint data, the manufacturer identified that a previous manufacturing process update may, under certain conditions, contribute to overpressure alerts. In response, the Instructions for Use (IFU) have been clarified to ensure proper device tightening., In the event of an overpressure condition, AirSeal® is designed to respond automatically via a built-in venting system, which activates when the pressure exceeds...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-06-11
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/cf939f06-deb7-4ed8-b7ab-dd78d6c1aab8