Bard Access Systems, Inc.: Needle, Hypodermic, Single Lumen

Bard Access Systems is recalling five models of BD Needle Kits for Powered Driver used to place intraosseous needles (emergency access to bone marrow when veins cannot be accessed) because the obturator may lock to the needle hub and become difficult or impossible to remove due to manufacturing tolerances and unclear instructions. This locking issue may require placement of a new needle and delay emergency treatment. The recall affects products distributed across most U.S. states and in Canada and Singapore.

Recalling firm
Bard Access Systems, Inc.
Product
BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK
Reason
After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-07-30
Status
Open, Classified
Areas affected
AL, AZ, AR, CA, CO, FL, GA, ID, IL, IN, IA, KS, ...