Philips Medical Systems Nederland B.v.: Allura System; System Code Description (model Numbers): Allura Xper Fd10 (722003, 722010, 722026), Allura Xper Fd10 Or Table (722022, 722033), Allu...

Recall Summary Philips is recalling multiple models of Allura medical imaging systems worldwide because the hard drives in their computers may degrade after six years of use, potentially causing loss of imaging capability, motorized movement, or data. The affected systems are used in medical facilities across the US and numerous countries globally. This is classified as a Class II recall, meaning the problem could cause serious health consequences but is not immediately life-threatening.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper FD20 (722006, 722012, 722028), Allura Xper FD20 Biplane (722008, 722013), Allura Xper FD20 OR Table (722015, 722023, 722035), Allura Xper FD20 Biplane OR Table (722025), Allura Xper FD20/10 (722029), Allura Xper FD20/15 (722058), Allura Xper FD20/15 OR Table (722059), Allura Xper FD20/20 (722038), Allura Xper FD20/20 OR Table (722039);
Reason
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-04-30
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussal, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus,...

Official recall notice from U.S. Food and Drug Administration

Recall details on Nibble Recalls: https://trynibble.app/recalls/cd9168a2-0034-42e8-9c75-8e1cb627776d