Fore-sight Elite Module Tissue Oximeter

Recalling firm
Edwards Lifesciences (New Zealand) Limited
Product
Fore-Sight Elite Module Tissue Oximeter. [Device]. Model: Catalog No. HEMFSM10. Affected: Serial numbers: 2008042, 2008043, 2008045, 2008046, 2008047, 21120094, 21120098, 21120099, 21120100, 21120692
Reason
The manufacturer has identified that, during manufacturing, the laterality identification labels may have been omitted or incorrectly applied to the device., Specifically, there is a possibility that the labels on the green cables were switched, resulting in the channel 1 cable being labelled as channel 2 and the channel 2 cable being labelled as channel 1. This may cause a discrepancy between the left and right values displayed on the monitor and the corresponding labels on the cables.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-28
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE